Citation: Ying Ying Zhong, Wen Fang Zhou, Xue Ling Zeng, Ming Li Ye, Yan Zhu. Trace analysis of anions in organic matrices by ion chromatography coupled with a novel reversed-phase column for on-line sample pretreatment[J]. Chinese Chemical Letters, 2011, 22(4): 461-464. doi: 10.1016/j.cclet.2010.11.011
Citation: Ying Ying Zhong, Wen Fang Zhou, Xue Ling Zeng, Ming Li Ye, Yan Zhu. Trace analysis of anions in organic matrices by ion chromatography coupled with a novel reversed-phase column for on-line sample pretreatment[J]. Chinese Chemical Letters, 2011, 22(4): 461-464. doi: 10.1016/j.cclet.2010.11.011
Trace analysis of anions in organic matrices by ion chromatography coupled with a novel reversed-phase column for on-line sample pretreatment
摘要:
An ion chromatography (IC) system coupled with on-line column-switching technique was used to determine anions of μg/g levels in organic chemicals of analytical reagent grade. A novel polystyrene-divinylbenzene-carbon nanotube (PS-DVB-CNT) stationary phase was utilized for matrix elimination. A calibration study was conducted by preparing and analyzing eight concentrations (between 10 and 5000 μg/L) of eight standards in deionized water. The linearity was between 0.9978 and 1. And the detection limits ranged from 1.54 μg/L to 10.02 μg/L. A spiking study was performed on two representative organic chemicals. The recoveries were between 84.3% and 119.6%.
English
Trace analysis of anions in organic matrices by ion chromatography coupled with a novel reversed-phase column for on-line sample pretreatment
Abstract:
An ion chromatography (IC) system coupled with on-line column-switching technique was used to determine anions of μg/g levels in organic chemicals of analytical reagent grade. A novel polystyrene-divinylbenzene-carbon nanotube (PS-DVB-CNT) stationary phase was utilized for matrix elimination. A calibration study was conducted by preparing and analyzing eight concentrations (between 10 and 5000 μg/L) of eight standards in deionized water. The linearity was between 0.9978 and 1. And the detection limits ranged from 1.54 μg/L to 10.02 μg/L. A spiking study was performed on two representative organic chemicals. The recoveries were between 84.3% and 119.6%.
-
Key words:
- Ion chromatography
- / Column switching
- / Trace anions
- / Organic matrix
- / Polymer stationary phase